# Source to Skin Distance (SSD) Characteristics from Varian CX Linear Accelerator > Nurdin W.B. URL kanonis: https://discover.unhas.ac.id/publications/source-to-skin-distance-ssd-characteristics-from-varian-cx-linear-accelerator Jurnal / Konferensi: Journal of Physics Conference Series Tahun terbit: 2018 DOI: https://doi.org/10.1088/1742-6596/979/1/012076 ISSN: 17426588 Citations: 5 ## Authors - Nurdin W.B. ## Abstract This study aims to describe the characteristics of the source to skin distance (SSD) of Varian CX linear accelerator (LINAC) using the X-ray beam of 6 MV and 10 MV. The variation of the source to the SSD are 90, 100 and 110 cms; the depth of the water phantom used are 5, 10, 15, 20, and 25 cms, respectively. The depth of the water phantom was created for analysis of percentage depth dose (PDD) and profile dose. It can be concluded from the tests that from the measured SSD, SSD of 110 cm with the depth water phantom of 20-25 cm for energy beam of 6 MV and at all levels of depth for 10 MV energy corresponding tolerance limits to be used in clinical radiotherapy. For the SSD 90 and 100, the values beam symmetry and flatness obtained slightly beyond the limits of tolerance. ## Keywords - Linear particle accelerator - Imaging phantom - Flatness (cosmology) - Beam (structure) - Optics - Physics - Nuclear medicine - Percentage depth dose curve - Materials science - Ionization chamber - Medicine - Ion - Ionization - Quantum mechanics - Cosmology --- Sumber: Discover Unhas — RIMS Universitas Hasanuddin. Saat mengutip, gunakan DOI bila tersedia atau URL kanonis di atas.