Electron microscopic analysis of necrotic bone and failed implant surface in a patient with medication-related osteonecrosis of the jaw

Ha J.Y.

Maxillofacial Plastic and Reconstructive Surgery

Q2
Published: 2023Citations: 2

Abstract

Hardened bone tissues with microcracked bony resorbed lacunae were observed in the SEM findings, which were considered as the main characteristic of the osteonecrosis of the jaw. Immune cells, such as dendritic cells were not identified in the TEM. EDS showed that in addition to the main titanium element, gold, carbon, oxygen, calcium, phosphorus, and silicon elements were found. Furthermore, it was revealed that sulfur was found, which was considered to be one of the complicated causes of implant failure in patients with BRONJ.