# Measurement of x-ray beam quality and patient dose information from analog mammography > Fitriani URL kanonis: https://discover.unhas.ac.id/publications/measurement-of-x-ray-beam-quality-and-patient-dose-information-from-analog-mammo Jurnal / Konferensi: Iop Conference Series Materials Science and Engineering Tahun terbit: 2019 DOI: https://doi.org/10.1088/1757-899X/546/3/032009 ISSN: 17578981 Citations: 1 ## Authors - Fitriani ## Abstract Measurement of X-ray beam quality and patient dose information from analog mammography have been studied by using an X-ray multimeter from 2016-2018. Parameters in the study are voltage accuracy, reproducibility test, radiation output linearity, radiation beam quality test, and patient dose information. The patient dose information using PMMA phantom with a thickness 4 cm in the Mean Glandular Dose (MGD) obtained dose values in 2016, 2017, 2018 are 0.982 mGy, 0.982 mGy, 0.953mGy, respectively. This study aims to determine the feasibility of mammography, while patient doses are compared with the standars set by the IAEA. The results showed that the suitability test of X-ray beam quality and patient dose information still comply the requirements recommended by the IAEA. ## Keywords - Mammography - Imaging phantom - Reproducibility - Nuclear medicine - Medicine - Quality assurance - Laser beam quality - Radiation dose - Multimeter - Medical physics - Beam (structure) - Dosimetry - Digital mammography - External beam radiation - Thermoluminescent dosimeter - Materials science - Radiation therapy - Optics - Radiology - Voltage - Dosimeter - Physics - Mathematics - Statistics - Laser beams - Cancer - Breast cancer - External quality assessment - Internal medicine - Pathology - Laser - Quantum mechanics --- Sumber: Discover Unhas — RIMS Universitas Hasanuddin. Saat mengutip, gunakan DOI bila tersedia atau URL kanonis di atas.