Measurement of x-ray beam quality and patient dose information from analog mammography

Fitriani

Iop Conference Series Materials Science and Engineering

Published: 2019Citations: 1

Abstract

Measurement of X-ray beam quality and patient dose information from analog mammography have been studied by using an X-ray multimeter from 2016-2018. Parameters in the study are voltage accuracy, reproducibility test, radiation output linearity, radiation beam quality test, and patient dose information. The patient dose information using PMMA phantom with a thickness 4 cm in the Mean Glandular Dose (MGD) obtained dose values in 2016, 2017, 2018 are 0.982 mGy, 0.982 mGy, 0.953mGy, respectively. This study aims to determine the feasibility of mammography, while patient doses are compared with the standars set by the IAEA. The results showed that the suitability test of X-ray beam quality and patient dose information still comply the requirements recommended by the IAEA.

Other files and links

Fingerprint

MammographySciences
Imaging phantomSciences
ReproducibilitySciences
Nuclear medicineSciences
MedicineSciences
Quality assuranceSciences
Laser beam qualitySciences
Radiation doseSciences
MultimeterSciences
Medical physicsSciences
Beam (structure)Sciences
DosimetrySciences
Digital mammographySciences
External beam radiationSciences
Thermoluminescent dosimeterSciences
Materials scienceSciences
Radiation therapySciences
OpticsSciences
RadiologySciences
VoltageSciences
DosimeterSciences
PhysicsSciences
MathematicsSciences
StatisticsSciences
Laser beamsSciences
CancerSciences
Breast cancerSciences
External quality assessmentSciences
Internal medicineSciences
PathologySciences
LaserSciences
Quantum mechanicsSciences