# Deviation Value for Conventional X-ray in Hospitals in South Sulawesi Province from 2014 to 2016 > Bachtiar I. URL kanonis: https://discover.unhas.ac.id/publications/deviation-value-for-conventional-x-ray-in-hospitals-in-south-sulawesi-province-f Jurnal / Konferensi: Journal of Physics Conference Series Tahun terbit: 2018 DOI: https://doi.org/10.1088/1742-6596/979/1/012081 ISSN: 17426588 Citations: 1 ## Authors - Bachtiar I. ## Abstract This paper describes the conventional X-ray machine parameters tested in the region of South Sulawesi from 2014 to 2016. The objective of this research is to know deviation of every parameter of conventional X-ray machine. The testing parameters were analyzed by using quantitative methods with participatory observational approach. Data collection was performed by testing the output of conventional X-ray plane using non-invasive x-ray multimeter. The test parameters include tube voltage (kV) accuracy, radiation output linearity, reproducibility and radiation beam value (HVL) quality. The results of the analysis show four conventional X-ray test parameters have varying deviation spans, where the tube voltage (kV) accuracy has an average value of 4.12%, the average radiation output linearity is 4.47% of the average reproducibility of 0.62% and the averaged of the radiation beam (HVL) is 3.00 mm. ## Keywords - Reproducibility - Multimeter - Standard deviation - Beam (structure) - Voltage - Linearity - Radiation - Optics - Statistics - Materials science - Physics - Mathematics - Electrical engineering - Engineering --- Sumber: Discover Unhas — RIMS Universitas Hasanuddin. Saat mengutip, gunakan DOI bila tersedia atau URL kanonis di atas.